Maquet Critical Care, MCC/22/008, Servo-u, Servo-n, Servo-u MR and Servo-air

 

 
 
 
Maquet Critical Care, MCC/22/008, Servo-u, Servo-n, Servo-u MR and Servo-air
It has been reported in complaints that ventilators have generated a combination of alarms; Technical Error 10 (Ventilation stopped), Technical Error 16 (Ventilation error) and Technical Error 55 (Communication errors) and stopped ventilating.



 
 
 
Maquet Critical Care, MCC/22/007, Servo-air
During new environmental tests, which includes the possibility that the user steps on the wheelbase, the earlier passed test results for Servo-air regarding the standard for vertical forces (IEC 60601-1 §9.4.2.3 b) could not be reproduced when testing an equivalent product.



 
 
 