RaySearch Laboratories AB (publ), FSN-65340, RayCare 2C SP2, 3A SP1, 3B and 4A

 

 
 
RaySearch Laboratories AB (publ), FSN-65340, RayCare 2C SP2, 3A SP1, 3B and 4A
This notice concerns an issue found in the Image viewer of RayCare PACS when visualizing a fusion with a study consisting of multiple image series in the same frame of reference.



 
 
Datascope Corporation, FSN-2248146-05/29/2020-001-R, Datascope Intra-Aortic Ballon Catheters (IABs)
Datascope/Getinge is initiating a voluntary Recall-Removal involving certain Intra-Aortic Balloon Catheters (IABs) that may not meet the requirement for endotoxin per AAMI ST72.



 
 
William Cook Europe ApS, FSN-2020FA0005, Zenith Alpha® Abdominal Endovascular Graft
Cook Medical has identified that specific lots of the Zenith Alpha® Abdominal Endovascular Graft (21MB main body) may contain excess glue within the handle of the delivery system, resulting in an inability to fully deploy the graft via the standard method or troubleshooting deployment method provided in the IFU.



 
 