vrijdag 9 augustus 2019

IGJ waarschuwingen: Hamilton Medical Nederland CER39384- HAMILTON-C3

Hamilton Medical Nederland CER39384- HAMILTON-C3
The analysis of a customer complaint has identified an issue related to SpO2 monitoring with the use of the HAMILTON-C3 ventilator. When measuring SpO2, the software of the HAMILTON-C3 can run into a system or CPU overload causing a screen freeze. The alarm buzzer sounds and the device needs to be reset.